5C1e Curved Array

5C1e Curved Array b Footnote:
*Cleaning is an important first step for all transducer types. Every transducer used in a critical procedure must be sterilized as the final reprocessing step.

9085/ View Product Page

Applicable Documents:

Print Table

Cleaning Agent

Compatibility

Compatibility Type

Comments

Cleaning d Footnote:
b) Transducer connector is not immersible

3E-Zyme (Manual Cleaning)

Yes

Validated

3E-Zyme (Automated use)

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Bodedex forte

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

CIDEZYME XTRA Multi-Enzymatic Detergent/ CIDEZYME GL Enzymatic Detergent

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Gigazyme

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Intercept Wipes/Intercept Detergent

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Korsolex Endo-Cleaner 0,5%

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

MATRIX Biofilm Remover

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Neodisher MediClean Forte

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

pH neutral (pH 6-8), non-corrosive cleaning products intended for medical devices

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Prolystica 2x concentration. Enzymatic

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Revital-Ox Bedside Complete/2X Concentrate Enzymatic Detergent/Enzymatic Detergents

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Sekusept MultiEnzyme P

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Suma Med Enzyme

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Disinfection d Footnote:
b) Transducer connector is not immersible

Korsolex Basic (Manual Disinfection)

Yes

Validated

Revital-Ox Resert/Resert XL HLD

Yes

Validated

Manual Disinfection

Ethanol 70% (wiping)

Yes

Validated

Manual Disinfection

Accel Prevention (wipes, ready-to-use liquid, concentrate)

No

Adaspor Single Shot

No

Antigermix S1

No

Astra VR (with approved disinfectant)

No

Bomix Plus

No

Cavi Wipes/CaviCide

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Cidex OPA

No

Cleanisept Wipes Forte

No

Gigasept FF

No

Glutaraldehyde 2% – 3.4%

No

Incidin OxyFoam/Incidin OxyWipe S

No

Isopropanol 70% (wiping)

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Korsolex Basic (Automated use)

No

Korsolex Endo Disinfectant 1%/Korsolex Extra

No

Medivators Advantage Plus Intercept (detergent), Rapicide PA Disinfectant flush: 70% isopropyl alcohol

No

Meliseptol Foam

No

Meliseptol Wipes Sensitive

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Metricide OPA Plus

No

Mikrobac Tissues

No

Mikrozoid AF Liquid

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Neodisher Endo Sept GA

No

Neodisher Endo SEPT PAC/Neodisher Septo PAC

No

Neodisher Septo DN

No

Nu-Cidex

No

OPAL

No

Rapicide/Rapicide OPA/28

No

Rely+On Perasafe

No

Revital-Ox Resert / Resert XL HLD (Automated use)

No

Sani Cloth Plus Wipes/Sani Cloth Super Wipes

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Sekusept Aktiv

No

Steranios 2%, 2% N.G., 2% E.C.S

No

Thermosept PAA

No

Tristel Duo for Ultrasound/Tristel Trio Wipe System

No

Tristel Fuse for Instruments (Tristel Fuse for Stella)

No

Trophon EPR / Trophon 2

No

Fit transducer completely above line mark.

Sterilization

Matachana 130LF, Webeco FA90, Webeco FA95 (60°C cycle)

No

RENO-20/RENO-30/RENO-D50: ECO Cycle

No

RENO-S90, RENO-S130, RENO-S130D: ECO Cycle

No

RENO-S90, RENO-S130, RENO-S130D: Non Lumen Cycle

No

STERIS System 1, 1E, 1 Plus and 1 Express

No

STERIS V-Pro 1 Plus, V-Pro 60Non lumen cycle

No

STERIS V-Pro maXNon lumen cycle or Flexible cycle

No

Sterrad 100 NX Express Cycle

No

Sterrad 100SOne cycle only (USA),Short cycle (rest of the world)

No

Sterrad 200Short cycle

No

Sterrad NX & 100NXStandard cycle

No

View Footnotes
Footnotes:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.
*Cleaning is an important first step for all transducer types. Every transducer used in a critical procedure must be sterilized as the final reprocessing step.
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.
b) Transducer connector is not immersible
c) Follow product manufacturer's instructions and do not exceed transducer-specified limits. See Care and Cleaning Guide for more information.
d) STERIS SYSTEM 1, SYSTEM 1 PLUS, and SYSTEM 1 Express have not been market cleared by the FDA.
e) Transducer has not been licensed by Health Canada