Print Table

Cleaning Agent

Compatibility

Compatibility Type

Comments

Cleaning d Footnote:
b) Transducer connector is not immersible

3E-Zyme (Manual Cleaning)

Yes

Validated

3E-Zyme (Automated use)

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Bodedex forte

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

CIDEZYME XTRA Multi-Enzymatic Detergent/ CIDEZYME GL Enzymatic Detergent

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Gigazyme

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Intercept Wipes/Intercept Detergent

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Korsolex Endo-Cleaner 0,5%

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

MATRIX Biofilm Remover

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Neodisher MediClean Forte

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

pH neutral (pH 6-8), non-corrosive cleaning products intended for medical devices

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Prolystica 2x concentration. Enzymatic

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Revital-Ox Bedside Complete/2X Concentrate Enzymatic Detergent/Enzymatic Detergents

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Sekusept MultiEnzyme P

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Suma Med Enzyme

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Disinfection d Footnote:
b) Transducer connector is not immersible

Medivators Advantage Plus Intercept (detergent), Rapicide PA Disinfectant flush: 70% isopropyl alcohol

Yes

Validated

Automated disinfection

Korsolex Basic (Manual Disinfection)

Yes

Validated

Tristel Fuse for Instruments (Tristel Fuse for Stella)

Yes

Validated

Manual Disinfection

Revital-Ox Resert/Resert XL HLD

Yes

Validated

Manual Disinfection

Ethanol 70% (wiping)

Yes

Validated

Manual Disinfection

Accel Prevention (wipes, ready-to-use liquid, concentrate)

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Adaspor Single Shot

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Antigermix S1

No

Astra VR (with approved disinfectant)

No

Bomix Plus

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Cavi Wipes/CaviCide

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Cidex OPA

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Cleanisept Wipes Forte

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Gigasept FF

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Glutaraldehyde 2% – 3.4%

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Incidin OxyFoam/Incidin OxyWipe S

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Isopropanol 70% (wiping)

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Korsolex Basic (Automated use)

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Korsolex Endo Disinfectant 1%/Korsolex Extra

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Meliseptol Foam

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Meliseptol Wipes Sensitive

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Metricide OPA Plus

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Mikrobac Tissues

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Mikrozoid AF Liquid

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Neodisher Endo Sept GA

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Neodisher Endo SEPT PAC/Neodisher Septo PAC

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Neodisher Septo DN

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Nu-Cidex

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

OPAL

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Rapicide/Rapicide OPA/28

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Rely+On Perasafe

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Revital-Ox Resert / Resert XL HLD (Automated use)

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Sani Cloth Plus Wipes/Sani Cloth Super Wipes

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Sekusept Aktiv

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Steranios 2%, 2% N.G., 2% E.C.S

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Thermosept PAA

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Tristel Duo for Ultrasound/Tristel Trio Wipe System

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Trophon EPR / Trophon 2

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Fit transducer completely above line mark.

Sterilization

STERIS System 1, 1E, 1 Plus and 1 Express c Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.
, d Footnote:
b) Transducer connector is not immersible

Yes

Validated

STERIS V-Pro 1 Plus, V-Pro 60Non lumen cycle c Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

Yes

Validated

STERIS V-Pro maXNon lumen cycle or Flexible cycle c Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

Yes

Validated

Sterrad NX & 100NXStandard cycle c Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

Yes

Validated

Sterrad 100 NX Express Cycle

Yes

Validated

Sterrad 100SOne cycle only (USA),Short cycle (rest of the world) c Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

Yes

Validated

Sterrad 200Short cycle c Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

Yes

Validated

Matachana 130LF, Webeco FA90, Webeco FA95 (60°C cycle)

No

RENO-20/RENO-30/RENO-D50: ECO Cycle

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

RENO-S90, RENO-S130, RENO-S130D: ECO Cycle

No

RENO-S90, RENO-S130, RENO-S130D: Non Lumen Cycle

Yes

Material Compatible

a Footnote:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

View Footnotes
Footnotes:
*Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.
*Cleaning is an important first step for all transducer types. Every transducer used in a critical procedure must be sterilized as the final reprocessing step.
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.
b) Transducer connector is not immersible
c) Follow product manufacturer's instructions and do not exceed transducer-specified limits. See Care and Cleaning Guide for more information.
d) STERIS SYSTEM 1, SYSTEM 1 PLUS, and SYSTEM 1 Express have not been market cleared by the FDA.
e) Transducer has not been licensed by Health Canada