Cleaning Agent
Compatibility
Compatibility Type
Comments
3E-Zyme (Manual Cleaning)
Yes
Validated
—
3E-Zyme (Automated use)
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—
Bodedex forte
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—
CIDEZYME XTRA Multi-Enzymatic Detergent/ CIDEZYME GL Enzymatic Detergent
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—
Gigazyme
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—
Intercept Wipes/Intercept Detergent
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—
Korsolex Endo-Cleaner 0,5%
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—
MATRIX Biofilm Remover
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—
Neodisher MediClean Forte
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—
pH neutral (pH 6-8), non-corrosive cleaning products intended for medical devices
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—
Prolystica 2x concentration. Enzymatic
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—
Revital-Ox Bedside Complete/2X Concentrate Enzymatic Detergent/Enzymatic Detergents
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—
Sekusept MultiEnzyme P
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—
Suma Med Enzyme
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—
Medivators Advantage Plus Intercept (detergent), Rapicide PA Disinfectant flush: 70% isopropyl alcohol
Yes
Validated
Automated disinfection
Korsolex Basic (Manual Disinfection)
Yes
Validated
—
Tristel Fuse for Instruments (Tristel Fuse for Stella)
Yes
Validated
Manual Disinfection
Revital-Ox Resert/Resert XL HLD
Yes
Validated
Manual Disinfection
Ethanol 70% (wiping)
Yes
Validated
Manual Disinfection
Accel Prevention (wipes, ready-to-use liquid, concentrate)
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—
Adaspor Single Shot
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—
Antigermix S1
No
—
—
Astra VR (with approved disinfectant)
No
—
—
Bomix Plus
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—
Cavi Wipes/CaviCide
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—
Cidex OPA
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—
Cleanisept Wipes Forte
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—
Gigasept FF
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—
Glutaraldehyde 2% – 3.4%
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—
Incidin OxyFoam/Incidin OxyWipe S
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—
Isopropanol 70% (wiping)
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—
Korsolex Basic (Automated use)
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—
Korsolex Endo Disinfectant 1%/Korsolex Extra
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—
Meliseptol Foam
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—
Meliseptol Wipes Sensitive
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—
Metricide OPA Plus
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—
Mikrobac Tissues
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—
Mikrozoid AF Liquid
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—
Neodisher Endo Sept GA
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—
Neodisher Endo SEPT PAC/Neodisher Septo PAC
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—
Neodisher Septo DN
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—
Nu-Cidex
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—
OPAL
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—
Rapicide/Rapicide OPA/28
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—
Rely+On Perasafe
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—
Revital-Ox Resert / Resert XL HLD (Automated use)
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—
Sani Cloth Plus Wipes/Sani Cloth Super Wipes
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—
Sekusept Aktiv
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—
Steranios 2%, 2% N.G., 2% E.C.S
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—
Thermosept PAA
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—
Tristel Duo for Ultrasound/Tristel Trio Wipe System
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—
Trophon EPR / Trophon 2
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.Fit transducer completely above line mark.
STERIS System 1, 1E, 1 Plus and 1 Express c Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer. , d Footnote: b) Transducer connector is not immersible
Yes
Validated
—
STERIS V-Pro 1 Plus, V-Pro 60Non lumen cycle c Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.
Yes
Validated
—
STERIS V-Pro maXNon lumen cycle or Flexible cycle c Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.
Yes
Validated
—
Sterrad NX & 100NXStandard cycle c Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.
Yes
Validated
—
Sterrad 100 NX Express Cycle
Yes
Validated
—
Sterrad 100SOne cycle only (USA),Short cycle (rest of the world) c Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.
Yes
Validated
—
Sterrad 200Short cycle c Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.
Yes
Validated
—
Matachana 130LF, Webeco FA90, Webeco FA95 (60°C cycle)
No
—
—
RENO-20/RENO-30/RENO-D50: ECO Cycle
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—
RENO-S90, RENO-S130, RENO-S130D: ECO Cycle
No
—
—
RENO-S90, RENO-S130, RENO-S130D: Non Lumen Cycle
Yes
Material Compatible
a Footnote: *Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.—