N13C5 (9062) Craniotomy Transducer

N13C5 (9062) Craniotomy Transducer

9062/ View Product Page

Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.

Follow product manufacturer’s instructions and do not exceed transducer-specified limits. See Care and Cleaning Guide for more information.

 

Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.
Follow product manufacturer’s instructions and do not exceed transducer specified limits. See Care and Cleaning Guide for more information.

Applicable Documents:

Print Table

Cleaning Agent

Compatibility

Compatibility Type

Comments

Cleaning c Footnote:
c) Cleaning is an important first step for all transducer types. Every transducer used in a critical procedure must be sterilized as the final reprocessing step.

3E-Zyme (Manual Cleaning)

Yes

Validated

3E-Zyme (Automated use)

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

Bodedex forte

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

CIDEZYME XTRA Multi-Enzymatic Detergent/ CIDEZYME GL Enzymatic Detergent

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

Gigazyme

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

Intercept Wipes/Intercept Detergent

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

Korsolex Endo-Cleaner 0,5%

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

MATRIX Biofilm Remover

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

Neodisher MediClean Forte

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

pH neutral (pH 6-8), non-corrosive cleaning products intended for medical devices

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

Prolystica 2x concentration. Enzymatic

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

Revital-Ox Bedside Complete/2X Concentrate Enzymatic Detergent/Enzymatic Detergents

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

Sekusept MultiEnzyme P

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

Suma Med Enzyme

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

Disinfection c Footnote:
c) Cleaning is an important first step for all transducer types. Every transducer used in a critical procedure must be sterilized as the final reprocessing step.

Medivators Advantage Plus Intercept (detergent), Rapicide PA Disinfectant flush: 70% isopropyl alcohol

Yes

Validated

Use of reprocessing method with the N13C5 is pending CE-marking.

Korsolex Basic (Manual Disinfection)

Yes

Validated

Manual Disinfection. Use of reprocessing method with the N13C5 is pending CE-marking.

Tristel Fuse for Instruments (Tristel Fuse for Stella)

Yes

Validated

Manual Disinfection. Use of reprocessing method with the N13C5 is pending CE-marking.

Revital-Ox Resert/Resert XL HLD

Yes

Validated

Manual Disinfection. Use of reprocessing method with the N13C5 is pending CE-marking.

Ethanol 70% (wiping)

Yes

Validated

Manual Disinfection

Accel Prevention (wipes, ready-to-use liquid, concentrate)

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

Adaspor Single Shot

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

Antigermix S1

No

Astra VR (with approved disinfectant)

No

Bomix Plus

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

Cavi Wipes/CaviCide

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

Cidex OPA

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

Cleanisept Wipes Forte

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

Gigasept FF

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

Glutaraldehyde 2% – 3.4%

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

Incidin OxyFoam/Incidin OxyWipe S

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

Isopropanol 70% (wiping)

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

Korsolex Basic (Automated use)

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

Korsolex Endo Disinfectant 1%/Korsolex Extra

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

Meliseptol Foam

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

Meliseptol Wipes Sensitive

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

Metricide OPA Plus

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

Mikrobac Tissues

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

Mikrozoid AF Liquid

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

Neodisher Endo Sept GA

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

Neodisher Endo SEPT PAC/Neodisher Septo PAC

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

Neodisher Septo DN

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

Nu-Cidex

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

OPAL

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

Rapicide/Rapicide OPA/28

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

Rely+On Perasafe

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

Revital-Ox Resert / Resert XL HLD (Automated use)

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

Sani Cloth Plus Wipes/Sani Cloth Super Wipes

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

Sekusept Aktiv

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

Steranios 2%, 2% N.G., 2% E.C.S

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

Thermosept PAA

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

Tristel Duo for Ultrasound/Tristel Trio Wipe System

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

Trophon EPR / Trophon 2

No

Sterilization

STERIS System 1, 1E, 1 Plus and 1 Express

Yes

Validated

STERIS SYSTEM 1, SYSTEM 1 PLUS, and SYSTEM 1 Express have not been market cleared by the FDA and are pending CE-mark for the N13C5.

STERIS V-Pro 1 Plus, V-Pro 60 Non lumen cycle

Yes

Validated

STERIS V-Pro maX Non lumen cycle or Flexible cycle

Yes

Validated

Sterrad NX and 100NX Standard cycle

Yes

Validated

Use of reprocessing method with the N13C5 is pending CE-marking.

Sterrad 100 NX Express Cycle

Yes

Validated

Use of reprocessing method with the N13C5 is pending CE-marking.

Sterrad 100S One cycle only (USA), Short cycle (rest of the world)

Yes

Validated

Use of reprocessing method with the N13C5 is pending CE-marking.

Sterrad 200 Short cycle

Yes

Validated

Use of reprocessing method with the N13C5 is pending CE-marking.

Matachana 130LF, Webeco FA90, Webeco FA95 (60°C cycle)

No

RENO-20/RENO-30/RENO-D50: ECO Cycle

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

RENO-S90, RENO-S130, RENO-S130D: ECO Cycle

No

RENO-S90, RENO-S130, RENO-S130D: Non Lumen Cycle

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.

View Footnotes
Footnotes:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method (when reprocessed according to the product/system IFU). Efficacy is not covered by this statement.
b) Transducer connector is not immersible
c) Cleaning is an important first step for all transducer types. Every transducer used in a critical procedure must be sterilized as the final reprocessing step.
d) Transducer has not been licensed by Health Canada