a magnifying glass
Intraoperative Biplane 8824

Intraoperative Biplane 8824

View Product Page

Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.
Follow product manufacturer’s instructions and do not exceed transducer specified limits. See Care and Cleaning Guide for more information.

Applicable Documents:

Print Table

Cleaning Agent

Compatibility

Compatibility Type

Comments

Cleaning c Footnote:
c) Cleaning is an important first step for all transducer types. Every transducer used in a critical procedure must be sterilized as the final reprocessing step.

3E-Zyme

Yes

Validated

Manual Cleaning

Yes

Material Compatible

a Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

Bodedex forte

Yes

Material Compatible

a Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

CIDEZYME XTRA Multi-Enzymatic Detergent/ CIDEZYME GL Enzymatic Detergent

Yes

Material Compatible

a Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

Gigazyme

Yes

Material Compatible

a Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

Intercept Wipes/Intercept Detergent

Yes

Material Compatible

a Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

Korsolex Endo-Cleaner 0,5%

Yes

Material Compatible

a Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

MATRIX Biofilm Remover

Yes

Material Compatible

a Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

Neodisher MediClean Forte

Yes

Material Compatible

a Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

pH neutral (pH 6-8), non-corrosive cleaning products intended for medical devices

Yes

Material Compatible

a Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

Prolystica 2x concentration. Enzymatic

Yes

Material Compatible

a Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

Revital-Ox Bedside Complete/2X Concentrate Enzymatic Detergent/Enzymatic Detergents

Yes

Material Compatible

a Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

Sekusept MultiEnzyme P

Yes

Material Compatible

a Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

Suma Med Enzyme

Yes

Material Compatible

a Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

Disinfection c Footnote:
c) Cleaning is an important first step for all transducer types. Every transducer used in a critical procedure must be sterilized as the final reprocessing step.

Medivators Advantage Plus; Intercept detergent, Rapicide PA disinfectant, flush: 70% isopropyl alcohol (Automated disinfection)

Yes

Validated

Automated Disinfection

Tristel Fuse for Instruments (Tristel Fuse for Stella)

Yes

Validated

Manual Disinfection

Revital-Ox Resert/Resert XL HLD

Yes

Validated

Manual Disinfection

Ethanol 70% (wiping)

Yes

Validated

Manual Disinfection

Korsolex Basic

Yes

Validated

Manual Disinfection

No

Yes

Material Compatible

a Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

20R3

Yes

Material Compatible

a Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

Accel Prevention (wipes, ready-to-use liquid, concentrate)

No

Adaspor Single Shot (Automated disinfection)

No

Antigermix S1 (Automated disinfection)

No

Astra VR

No

Bomix Plus

Yes

Material Compatible

a Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

Cavi Wipes/CaviCide

Yes

Material Compatible

a Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

Cidex OPA

Yes

Material Compatible

a Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

Cleanisept Wipes Forte

Yes

Material Compatible

a Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

Gigasept FF

Yes

Material Compatible

a Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

Glutaraldehyde 2% – 3.4%

Yes

Material Compatible

a Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

Incidin OxyFoam/Incidin OxyWipe S

Yes

Material Compatible

a Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

Isopropanol 70% (wiping)

Yes

Material Compatible

a Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

Korsolex Endo Disinfectant 1%/Korsolex Extra

Yes

Material Compatible

a Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

Meliseptol Foam

Yes

Material Compatible

a Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

Meliseptol Wipes Sensitive

Yes

Material Compatible

a Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

Metricide OPA Plus

Yes

Material Compatible

a Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

Mikrobac Tissues

Yes

Material Compatible

a Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

Mikrozoid AF Liquid

Yes

Material Compatible

a Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

Neodisher Endo Sept GA

Yes

Material Compatible

a Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

Neodisher Septo DN

Yes

Material Compatible

a Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

Nu-Cidex

Yes

Material Compatible

a Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

OPAL

Yes

Material Compatible

a Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

Rapicide

Yes

Material Compatible

a Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

RAPICIDE OPA/28

Yes

Material Compatible

a Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

Rely+On Perasafe

Yes

Material Compatible

a Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

Sani Cloth Plus Wipes/Super Wipes

Yes

Material Compatible

a Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

Sekusept Aktiv

No

Steranios 2%, 2% N.G., 2% E.C.S

Yes

Material Compatible

a Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

Thermosept PAA

No

Tristel Duo for Ultrasound

Yes

Material Compatible

a Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

Tristel Trio Wipe System

Yes

Material Compatible

a Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

Trophon EPR / Trophon 2

No

Sterilization

STERIS System 1, 1E, 1 Plus and 1 Express d Footnote:
d) STERIS SYSTEM 1, SYSTEM 1 PLUS, AND SYSTEM 1 EXPRESS are not market cleared by the FDA.

Yes

Validated

STERIS V-Pro 1 Plus, V-Pro 60 Non lumen cycle

Yes

Validated

STERIS V-Pro maX Non lumen cycle or Flexible cycle

Yes

Validated

STERIZONE VP4 c Footnote:
c) Cleaning is an important first step for all transducer types. Every transducer used in a critical procedure must be sterilized as the final reprocessing step.

Yes

Validated

Sterrad NX and 100NX Standard cycle

Yes

Validated

Valid for transducers with a serial number higher than 1910000. If in doubt, contact your BK representative.

Sterrad 100S One cycle only (USA), Short cycle (rest of the world)

Yes

Validated

Sterrad 200 Short cycle

Yes

Validated

Matachana 130LF, Webeco FA90, Webeco FA95 (60°C cycle)

Yes

Validated

Use a suitable connector lid and valve from the system manufacturer. See page 46.

No

RENO-20/RENO-30/RENO-D50: ECO Cycle

Yes

Material Compatible

a Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

RENO-S90, RENO-S130, RENO-S130D: ECO cycle and Non Lumen cycle

Yes

Material Compatible

a Footnote:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.

View Footnotes
Footnotes:
a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.
b) Transducer connector is not immersible.
c) Cleaning is an important first step for all transducer types. Every transducer used in a critical procedure must be sterilized as the final reprocessing step.
d) STERIS SYSTEM 1, SYSTEM 1 PLUS, AND SYSTEM 1 EXPRESS are not market cleared by the FDA.