T7P2m (9027) TEE Transducer

T7P2m (9027) TEE Transducer

9027/ View Product Page

Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.
Follow product manufacturer’s instructions and do not exceed transducer specified limits. See Care and Cleaning Guide for more information.

Applicable Documents:

Print Table
Tip until 100cm marker
- submersible
Handle and cable part
- not submersible, wiping only
Connector (and white flex relief )
- not submersible, wiping only

Validated Methods

3E-Zyme

CIVCO UltrOx HLD

Revital-Ox Resert XL HDL

Material Compatible Methods

Mild pH neutral detergent

Mild pH neutral detergent wipe

Mild pH neutral detergent wipe

Aniosyme DD1

Clinell Sporicidal wipes

Tristel Trio Wipes System

Cidex ADS, OPA and Nu-Cidex

Ethanol (EtOH) 70% wipes

Ethanol 70% wipes

Cidezyme / Enzol

Isopropyl alcohol (IPA) 70% wipes

IPA 70% wipes

Empower

Mikrozid® PAA wipes

Gigasept AF, Gigasept FF (Neu) and Gigasept PAA concentrate

Sani-Cloth AF3 wipes

Korsolex extra

Tristel Trio Wipes System

Metricide, Metricide 28, Metricide OPA Plus and Metricide Plus 30

Metrizyme

Neodisher MediClean forte/ Neodisher Septo 3000

Perasafe

Prolystica 2x conc.

Revital-Ox Resert XL HDL

Sekusept Aktiv

Steranios 2%, 2% N.G., 2% E.C.S

TD100 & TD5

Tristel Trio Wipe System

Wavicide 01

Warning: Extra care must be taken when cleaning this transducer, as there is no automatic cleaning system that can reprocess the entire transducer.