Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.
Follow product manufacturer’s instructions and do not exceed transducer specified limits. See Care and Cleaning Guide for more information.
Applicable Documents:
Cleaning Agent
Compatibility
Compatibility Type
Comments
3E-Zyme
Yes
Validated
Manual Cleaning
Yes
Material Compatible
a Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.—
Bodedex forte
Yes
Material Compatible
a Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.—
CIDEZYME XTRA Multi-Enzymatic Detergent/ CIDEZYME GL Enzymatic Detergent
Yes
Material Compatible
a Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.—
Gigazyme
Yes
Material Compatible
a Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.—
Intercept Wipes/Intercept Detergent
Yes
Material Compatible
a Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.—
Korsolex Endo-Cleaner 0,5%
Yes
Material Compatible
a Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.—
MATRIX Biofilm Remover
Yes
Material Compatible
a Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.—
Neodisher MediClean Forte
Yes
Material Compatible
a Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.—
pH neutral (pH 6-8), non-corrosive cleaning products intended for medical devices
Yes
Material Compatible
a Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.—
Prolystica 2x concentration. Enzymatic
Yes
Material Compatible
a Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.—
Revital-Ox Bedside Complete/2X Concentrate Enzymatic Detergent/Enzymatic Detergents
Yes
Material Compatible
a Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.—
Sekusept MultiEnzyme P
Yes
Material Compatible
a Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.—
Suma Med Enzyme
Yes
Material Compatible
a Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.—
Medivators Advantage Plus; Intercept detergent, Rapicide PA disinfectant, flush: 70% isopropyl alcohol (Automated disinfection)
Yes
Validated
Automated Disinfection
Revital-Ox Resert/Resert XL HLD
Yes
Validated
Manual Disinfection
Ethanol 70% (wiping)
Yes
Validated
Manual Disinfection
Korsolex Basic
Yes
Validated
Manual Disinfection
Yes
Material Compatible
a Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.—
Yes
Material Compatible
a Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.—
20R3
No
—
—
Accel Prevention (wipes, ready-to-use liquid, concentrate)
Yes
Material Compatible
a Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.—
Adaspor Single Shot (Automated disinfection)
Yes
Material Compatible
a Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.—
Antigermix S1 (Automated disinfection)
No
—
—
Astra VR
No
—
—
Bomix Plus
Yes
Material Compatible
a Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.—
Cavi Wipes/CaviCide
Yes
Material Compatible
a Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.—
Cidex OPA
Yes
Material Compatible
a Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.—
Cleanisept Wipes Forte
Yes
Material Compatible
a Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.—
Gigasept FF
Yes
Material Compatible
a Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.—
Glutaraldehyde 2% – 3.4%
Yes
Material Compatible
a Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.—
Incidin OxyFoam/Incidin OxyWipe S
Yes
Material Compatible
a Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.—
Isopropanol 70% (wiping)
Yes
Material Compatible
a Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.—
Korsolex Endo Disinfectant 1%/Korsolex Extra
No
—
—
Meliseptol Foam
Yes
Material Compatible
a Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.—
Meliseptol Wipes Sensitive
Yes
Material Compatible
a Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.—
Metricide OPA Plus
Yes
Material Compatible
a Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.—
Mikrobac Tissues
Yes
Material Compatible
a Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.—
Mikrozoid AF Liquid
Yes
Material Compatible
a Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.—
Neodisher Endo Sept GA
Yes
Material Compatible
a Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.—
Neodisher Septo DN
Yes
Material Compatible
a Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.—
Nu-Cidex
No
—
—
OPAL
Yes
Material Compatible
a Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.—
Rapicide
Yes
Material Compatible
a Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.—
RAPICIDE OPA/28
Yes
Material Compatible
a Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.—
Rely+On Perasafe
Yes
Material Compatible
a Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.—
Sani Cloth Plus Wipes/Super Wipes
Yes
Material Compatible
a Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.—
Sekusept Aktiv
Yes
Material Compatible
a Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.—
Steranios 2%, 2% N.G., 2% E.C.S
Yes
Material Compatible
a Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.—
Thermosept PAA
Yes
Material Compatible
a Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.—
Tristel Duo for Ultrasound
No
—
—
Tristel Fuse for Instruments (Tristel Fuse for Stella)
No
—
—
Tristel Trio Wipe System
No
—
—
Trophon EPR / Trophon 2
Yes
Material Compatible
a Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.Fit transducer completely above line mark.
STERIS System 1, 1E, 1 Plus and 1 Express d Footnote: d) STERIS SYSTEM 1, SYSTEM 1 PLUS, AND SYSTEM 1 EXPRESS are not market cleared by the FDA.
Yes
Validated
—
STERIS V-Pro 1 Plus, V-Pro 60 Non lumen cycle
Yes
Validated
—
STERIS V-Pro maX Non lumen cycle or Flexible cycle
Yes
Validated
—
Sterrad 100S One cycle only (USA), Short cycle (rest of the world)
Yes
Validated
—
Sterrad 200 Short cycle
Yes
Validated
—
No
—
—
Matachana 130LF, Webeco FA90, Webeco FA95 (60°C cycle)
No
—
—
RENO-20/RENO-30/RENO-D50: ECO Cycle
Yes
Material Compatible
a Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.—
RENO-S90, RENO-S130, RENO-S130D: ECO cycle and Non Lumen cycle
Yes
Material Compatible
a Footnote: a) This table of reprocessing methods only indicates that BK Medical has evaluated the transducers for material compatibility, when reprocessed according to the product/system IFU, and not for their efficacy in attaining the appropriate level of reprocessing. Ensure that the reprocessing chamber is an appropriate size for the transducer.—
STERIZONE VP4
No
—
—
Sterrad NX and 100NX Standard cycle
No
—
—