a magnifying glass
14L3e (9086) Linear Transducer

14L3e (9086) Linear Transducer b Footnote:
b) Transducer connector is not immersible.

9086/ View Product Page

Follow local regulations for minimum reprocessing. Check table 4 on page 29 of the Care and Cleaning Guide.
Follow product manufacturer’s instructions and do not exceed transducer specified limits. See Care and Cleaning Guide for more information.

Applicable Documents:

Print Table

Cleaning Agent

Compatibility

Compatibility Type

Comments

Cleaning c e Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method when reprocessed according to the product/system IFU. Efficacy is not covered by this statement.

3E-Zyme

Yes

Validated

14L3e (9086) Linear Transducer b Footnote:
b) Transducer connector is not immersible.

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method when reprocessed according to the product/system IFU. Efficacy is not covered by this statement.

Bodedex forte

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method when reprocessed according to the product/system IFU. Efficacy is not covered by this statement.

CIDEZYME XTRA Multi-Enzymatic Detergent/ CIDEZYME GL Enzymatic Detergent

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method when reprocessed according to the product/system IFU. Efficacy is not covered by this statement.

Gigazyme

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method when reprocessed according to the product/system IFU. Efficacy is not covered by this statement.

Intercept Wipes/Intercept Detergent

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method when reprocessed according to the product/system IFU. Efficacy is not covered by this statement.

Korsolex Endo-Cleaner 0,5%

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method when reprocessed according to the product/system IFU. Efficacy is not covered by this statement.

MATRIX Biofilm Remover

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method when reprocessed according to the product/system IFU. Efficacy is not covered by this statement.

Neodisher MediClean Forte

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method when reprocessed according to the product/system IFU. Efficacy is not covered by this statement.

pH neutral (pH 6-8), non-corrosive cleaning products intended for medical devices

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method when reprocessed according to the product/system IFU. Efficacy is not covered by this statement.

Prolystica 2x concentration. Enzymatic

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method when reprocessed according to the product/system IFU. Efficacy is not covered by this statement.

Revital-Ox Bedside Complete/2X Concentrate Enzymatic Detergent/Enzymatic Detergents

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method when reprocessed according to the product/system IFU. Efficacy is not covered by this statement.

Sekusept MultiEnzyme P

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method when reprocessed according to the product/system IFU. Efficacy is not covered by this statement.

Suma Med Enzyme

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method when reprocessed according to the product/system IFU. Efficacy is not covered by this statement.

Disinfection c e Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method when reprocessed according to the product/system IFU. Efficacy is not covered by this statement.

Korsolex Basic

Yes

Validated

Ethanol 70% (wiping)

Yes

Validated

Revital-Ox Resert/Resert XL HLD

Yes

Validated

Tristel Fuse for Instruments (Tristel Fuse for Stella)

Yes

Validated

CIVCO UltrOx HLD

Yes

Validated

14L3e (9086) Linear Transducer

No

14L3e (9086) Linear Transducer

No

Accel Prevention (wipes, ready-to-use liquid, concentrate)

No

Adaspor Single Shot (Automated disinfection)

No

Antigermix S1 (Automated disinfection)

No

Astra VR (with approved disinfectant)

No

Bomix Plus

No

Cavi Wipes/CaviCide

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method when reprocessed according to the product/system IFU. Efficacy is not covered by this statement.

Cidex OPA

No

Cleanisept Wipes Forte

No

Gigasept FF

No

Glutaraldehyde 2% – 3.4%

No

Incidin OxyFoam/Incidin OxyWipe S

No

Isopropanol 70% (wiping)

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method when reprocessed according to the product/system IFU. Efficacy is not covered by this statement.

Korsolex Endo Disinfectant 1%/Korsolex Extra

No

Medivators Advantage PlusIntercept (detergent), Rapicide PA Disinfectantflush: 70% isopropyl alcohol (Automated disinfection)

No

Meliseptol Foam

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method when reprocessed according to the product/system IFU. Efficacy is not covered by this statement.

Meliseptol Wipes Sensitive

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method when reprocessed according to the product/system IFU. Efficacy is not covered by this statement.

Metricide OPA Plus

No

Mikrobac Tissues

No

Mikrozoid AF Liquid

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method when reprocessed according to the product/system IFU. Efficacy is not covered by this statement.

Neodisher Endo Sept GA

No

Neodisher Endo SEPT PAC/Neodisher Septo PAC

No

Neodisher Septo DN

No

Nu-Cidex

No

OPAL

No

Rapicide/Rapicide OPA/28

No

Rely+On Perasafe

No

Sani Cloth Plus Wipes/Sani Cloth Super Wipes

Yes

Material Compatible

a Footnote:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method when reprocessed according to the product/system IFU. Efficacy is not covered by this statement.

Sekusept Aktiv

No

Steranios 2%, 2% N.G., 2% E.C.S

No

Thermosept PAA

No

Tristel Duo for Ultrasound/Tristel Trio Wipe System

No

Trophon EPR / Trophon 2

No

Fit transducer completely above line mark.

Sterilization f Footnote:
f) All sterilization methods are automated.

14L3e (9086) Linear Transducer

No

Matachana 130LF, Webeco FA90, Webeco FA95 (60°C cycle)

No

RENO-20/RENO-30/RENO-D50: ECO Cycle

No

RENO-S90, RENO-S130, RENO-S130D: ECO cycle and Non Lumen cycle

No

STERIS System 1, 1E, 1 Plus and 1 Express

No

STERIS V-Pro 1 Plus, V-Pro 60 Non lumen cycle

No

STERIS V-Pro maX Non lumen cycle or Flexible cycle

No

Sterrad 100 NX Express Cycle

No

Sterrad 100S One cycle only (USA), Short cycle (rest of the world)

No

Sterrad 200 Short cycle

No

Sterrad NX and 100NX Standard cycle

No

View Footnotes
Footnotes:
a) “Material compatible” indicates that BK Medical has evaluated the device’s material compatibility with the reprocessing method when reprocessed according to the product/system IFU. Efficacy is not covered by this statement.
b) Transducer connector is not immersible.
c) Cleaning is an important first step for all transducer types. Every transducer used in a critical procedure must be sterilized as the final reprocessing step.
d) Transducer has not been licensed by Health Canada.
e) All cleaning methods and all disinfection methods are manual unless otherwise noted.
f) All sterilization methods are automated.